Imaging®: doses investigation

The analysis of the Imaging component renders unique filter patterns that provide qualitative information about the conditions during exposure. Imaging to identify static, dynamic, or contamination conditions is automatically performed for all beta and low-energy photon measurements yielding a dose in excess of 5 mSv. Imaging capabilities are inconclusive at energies exceeding 150 keV.

Luxel+ filter pack

Luxel+ filter pack

The analysis of the Imaging component renders unique filter patterns that provide qualitative information about the conditions during exposure. Imaging to identify static, dynamic, or contamination conditions is automatically performed for all beta and low-energy photon measurements yielding a dose in excess of 5 mSv. Imaging capabilities are inconclusive at energies exceeding 150 keV.

Imaging is not available for doses less than 0.5 mSv.

 

A static exposure image indicates the dosemeter was still with no movement at the time of exposure. This is verified by the distinct grid patterns in the filter pack imaging area. A static exposure implies that the dosemeter may not have been worn at the time of exposure, and an accidental exposure may have occurred.

A dynamic exposure image indicates the dosemeter was moving at the time of exposure. This is verified by the blurred grid patterns in the filter pack imaging area. A dynamic exposure implies that the dosemeter was worn at the time of exposure, and the reported dose is valid.

Reanalysis or imaging at doses can be requested to our Customer Service.